dt-bindings: iio: temperature: add MLX90632 device bindings

Add device tree bindings for MLX90632 IR temperature sensor.

Signed-off-by: Crt Mori <cmo@melexis.com>
Reviewed-by: Rob Herring <robh@kernel.org>
Signed-off-by: Jonathan Cameron <Jonathan.Cameron@huawei.com>
This commit is contained in:
Crt Mori 2018-01-11 11:20:10 +01:00 committed by Jonathan Cameron
parent 47a3616348
commit a8062aa116
1 changed files with 28 additions and 0 deletions

View File

@ -0,0 +1,28 @@
* Melexis MLX90632 contactless Infra Red temperature sensor
Link to datasheet: https://www.melexis.com/en/documents/documentation/datasheets/datasheet-mlx90632
There are various applications for the Infra Red contactless temperature sensor
and MLX90632 is most suitable for consumer applications where measured object
temperature is in range between -20 to 200 degrees Celsius with relative error
of measurement below 1 degree Celsius in object temperature range for
industrial applications. Since it can operate and measure ambient temperature
in range of -20 to 85 degrees Celsius it is suitable also for outdoor use.
Be aware that electronics surrounding the sensor can increase ambient
temperature. MLX90632 can be calibrated to reduce the housing effect via
already existing EEPROM parameters.
Since measured object emissivity effects Infra Red energy emitted, emissivity
should be set before requesting the object temperature.
Required properties:
- compatible: should be "melexis,mlx90632"
- reg: the I2C address of the sensor (default 0x3a)
Example:
mlx90632@3a {
compatible = "melexis,mlx90632";
reg = <0x3a>;
};